Inductively Coupled Plasma Optical Emission Spectrometer ICP-6800
Main Features
Performance stability
The use of all-solid-state rf power supply, with small size, high efficiency, stable output power, with a variety of protection functions and many other advantages, the load using fully automatic matching technology, matching speed, improve the efficiency of the use of power supply and the stability of the instrument, and make the whole ignition process simple and convenient.
Injection automation
The instrument is equipped with a four-channel twelve-roller high-precision peristaltic pump, which can guarantee the precision of injection and prevent liquid accumulation at the same time. The speed of the peristaltic pump can be adjusted continuously to meet various test requirements of customers.
Flow control parameters
Adopt imported high precision mass flow controller.
High precision constant temperature system
The whole optical path of adopt the precision constant temperature protection system. The temperature control can be set in real time according to the actual ambient temperature of the customer's laboratory, without the need to turn on the air conditioner continuously for a long time. Temperature control precision ≤±0.1%, precision constant temperature protection system to ensure the stability of the optical path and makes the test data more stable.
RFPower technical parameter:
Circuit type :solid-state RF Power supply, with function of automatch |
Power Output :800W-1200W |
Power Output Stability :<0.3% |
Escaped RF radiation :30cm away from the instrument, |
electric fiekd: E<2V/m |
Frequency Stablity :<0.1% |
Frequercy :27.12MHz±0.05% |
Sampling System Technical Parameter
Output working coil imer diameter :25mm
Torpue tube :Three concentric, extemal diameter: 20mm
Coaxial nebulizer :Outer dameter 6mm
Double barrel atomizing chamber :Outer diameter 34mm
Gas Flow Controls
Plasma Argon Flowmeter :(100-1000)L/h(1.6-16L/min)
Auxiliary Argon Fiowmeter: (10-100)L/h(0.16-1.66L/min)
Auxiliary Carrier Argon Flowmeter :(10-100)L/h(0.16-1.66L/min)
Pressune Maintaining Value :(0-0.4MPa)
Cooling Water:Temperature :20-25ºC,Rate of Flow>5L/min,Hydraulic Pressure>0.1MPa
Spectrometers
1 | Optics :Czerny-Turner type |
2 | Focal length :1000 mm |
3 | Grating :Ion Beam Etching Holographic Grating, 3600L/mm or 2400L/mm |
4 | Reciprocal linear dispersion :0.26nm/mm |
5 | Resolution :≤0.007nm(3600 line grating); ≤0.015nm(2400 line grating) |
6 | Wavelength range :;3600 line grating:(190nm~500) nm; 2400 line grating:(190nm~800) nm |
7 | Minimum pace of stepping motor :≤0.0006 nm |
8 | Exit Slit :12μm |
9 | Entrance Slit :10μm |
Overall Performance
Wavelength range of scanning
195nm~500nm(3600L/mm line grating);195nm~800nm(2400L/mm line grating)
Repeatability
RSD(Relative Standard Deviation)≤1.5%
Stability
RSD(Relative Standard Deviation) ≤2%
Detection Limits (µg/L):
Our Company:
Exhibition:
Contact Us
Macylab Instruments Inc.
Macylab Instruments Inc.is a high-tech enterprise with independent intellectual property rights . Macy hold the pioneering concept "technology - because you change " , and as a business purpose, and constantly explore, bold innovation. Especially in the field of analytical instruments ,Macy constantly develop advanced products that enable Macy to become a high-quality instrument resource supplier and leader。
Macylab Instruments Inc.focus on UV/Vis spectrophotometer, including the 1100/1200 Series UV/Vis spectrophotometer, 1300 Series UV/Vis spectrophotometer, 1500 Series UV/Vis spectrophotometer, 1700 Series UV/Vis spectrophotometer meter, 1800 series double beam UV/Vis spectrophotometer and 1900 series double beam UV/Vis spectrophotometer.Our Atomic absorption spectrophotometer including : AA-1800C flame atomic absorption spectrophotometer, AA-1800H Flame / Graphite Furnace Intergrated atomic absorption spectrophotometer. At present, our products have been widely used in organic chemistry, inorganic chemistry, biochemistry, medicine, environmental protection, metallurgy, petroleum, agriculture and other fields. At the same time, aesthetic analysis using the rich experience accumulated in product mechanical structure, optical design, electrical application and software development, combined with the latest actual market demand, will soon launch a number of new analysis instrument.
Macylab's headquarter and production base is located in Shanghai, the marketing center in Beijing, and three R & D bases in Jiangsu, Shanghai, and Shandong. To take full advantage of the country's intellectual resources, MACY and the domestic and foreign research unit also conducted deep scientific research cooperation, will continue to put the research achievements into productivity .To better serve our customers, Macy has 12 domestic offices, tailored to meet the needs of your application solutions to improve the value of the products. At the same time continue to serve domestic users, Macy also with distribution in over 20 countries to establish a deep strategic partnership cooperation.
Along with Macylab accelerate the pace and rank among the global brand instruments, Macylab increasing demands of our own, and we also hope to get the care and support of the community, Let us work and outlook together.Technology, will change because you and us.